Product detail
DL-1060 Digital Multimeter
DL-1060 Series
on sale (Limited)
Singapore,Malaysia,
Thailand,Vietnam,
China,Korea Only
6 1/2 digit Dual display DMM
Specification
● High-speed sampling rate (50000 Rdgs/sec. at NPLC 0.001)
● Direct Thermocouple Measurement (Built-In Cold Junction to improve Accuracy)
● Dual Measurement & Dual Display (Including Thermocouple Measurement)
● Capacitance Measurement
● Several Measurements & Mathematic Functions
● Wide Range Current Measurement (Up to 10A)
● High Sensitivity(DCV: 0.1μV & Resistans: 100μΩ)
● Plug & Play Interface: Built-in USB (USBTMC)
● High Storage Memory (Up to 2000 Readings)
● Free Application Software
Rating
DC Characteristic Accuracy ±(% of reading * % pf range) *1
| Function |
Range *2 |
Resolution |
Input resistance |
1Year (23°C±5°C) |
| DC Voltage |
100.0000mV |
0.1μV |
10MΩ |
0.008 + 0.0045 |
| 1.000000V |
1μV |
10MΩ |
0.009 + 0.001 |
| 10.00000V |
10μV |
10MΩ |
0.012 + 0.002 |
| 100.0000V |
100μV |
10MΩ |
0.012 + 0.002 |
| 1000.000V |
1mV |
10MΩ |
0.02 + 0.00 |
| Function |
Range *2 |
Resolution |
Shunt resistance |
1Year (23°C±5°C) |
| DC Current |
10.00000mA |
10nA |
5.1Ω |
0.05 + 0.02 |
| 100.0000mA |
100nA |
5.1Ω |
0.05 + 0.01 |
| 1.000000A |
1 μA |
0.1Ω |
0.15 + 0.022 |
| 3.00000A |
10μA |
0.1Ω |
0.2 + 0.03 |
| 10.00000A |
10μA |
0.005Ω |
0.25 + 0.05 |
| Function |
Range |
Resolution |
Test Current |
1Year (23°C±5°C) |
| Resistance *3 |
100.0000 Ω |
100μΩ |
1mA |
0.02 + 0.005 |
| 1.000000 kΩ |
1mΩ |
1mA |
0.02 + 0.002 |
| 10.00000 kΩ |
10mΩ |
100μA |
0.02 + 0.002 |
| 100.0000 kΩ |
100mΩ |
10μA |
0.02 + 0.002 |
| 1.000000 MΩ |
1Ω |
1μA |
0.02 + 0.004 |
| 10.00000 MΩ |
10Ω |
0.1μA |
0.1 + 0.004 |
| 100.0000 MΩ |
100Ω |
0.1µA(parallel10MΩ) |
1.5 + 0.005 |
| Diode test |
1.0000V |
10μV |
1mA |
0.02 + 0.02 |
| Continuity |
1000.00Ω |
10mΩ |
1mA |
0.02 + 0.03 |
Frequency and Period Characteristics *4
| Function |
Range*2 |
Frequency Hz |
1 Year Aging |
| Frequency & PeriodFrequency |
100mV to 750V *4 |
10 to 40 |
0.03 |
| 40 to 300k |
0.02 |
FAC Characteristic Accuracy ±(% of reading + % of range) *1
| Function |
Range*2 |
Resolution |
FFrequency Hz |
1Year (23°C±5°C) |
| AC Voltage(TRMS) *5 |
100.0000mV |
0.1µV |
10 to 20k |
0.12 + 0.05 |
| 20k to 50k |
0.25 + 0.05 |
| 50k to 100k |
0.65 + 0.08 |
| 100k to 300k |
4.8 + 0.8 |
| 1.000000V to 750.000V *4 |
1µV to 1mV |
10 to 20k |
0.12 + 0.04 |
| 20k to 50k |
0.25 + 0.05 |
| 50k to 100k |
0.65 + 0.08 |
| 100k to 300k |
4.8 + 0.8 |
| AC Current(TRMS) *5 |
1.000000A |
1µV |
10 to 1k |
0.2 + 0.04 |
| 1k to 5k |
1.00 + 0.1 |
| 3.00000A |
10µV |
10 to 1k |
0.3 + 0.06 |
| 1k to 5k |
1.5 + 0.15 |
| 10.00000A |
10µV |
10 to 1k |
0.5 + 0.12 |
| 1k to 5k |
2.5 + 0.2 |
Capacitance Characteristic Accuracy ±(% of reading + % of range) *1
| Function |
Range |
Test Current |
1Year (23°C±5°C) |
| Capacitance *6 |
1nF |
10µA |
2.0 + 0.80 |
| 10nF |
10µA |
1.0 + 0.50 |
| 100nF |
100µA |
1.0 + 0.50 |
| 1µF |
100µA |
1.0 + 0.50 |
| 10µF |
100µA |
1.0 + 0.50 |
| 100µF |
1mA |
1.0 + 0.50 |
| 1mF |
1mA |
1.0 + 0.50 |
| 10mF |
1mA |
1.0 + 0.50 |
Thermocouple Characteristic *1
| Function |
Type |
Range |
1Year (23°C±5°C) |
| Thermocouple*7 |
1nF |
10µA |
2.0 + 0.80 |
| 10nF |
10µA |
1.0 + 0.50 |
| 100nF |
100µA |
1.0 + 0.50 |
| 1µF |
100µA |
1.0 + 0.50 |
| 10µF |
100µA |
1.0 + 0.50 |
| 100µF |
1mA |
1.0 + 0.50 |
| 1mF |
1mA |
1.0 + 0.50 |
| 10mF |
1mA |
1.0 + 0.50 |
General information
| Power supply |
100V / 120V / 220V/ 240V ± 10% |
| Power frequency |
50 / 60Hz ± 10% |
| Power consumption Max |
25VA |
| 0℃ to 50℃ |
25VA |
| Operating humidity |
0℃ to 31℃ 80% RH |
| Stock temperature |
-10℃ to 61℃ |
| Operating altitude |
Up to 2000m |
| Dimension |
214.6(W)×88.6(H)×280.7(D)mm |
| Weight |
Approx. 2.7kg |
| Safety regulation |
Approx. 2.7kg |
| EN61010-1:2001 (2nd Edition) |
Approx. 2.7kg |
| EMC regulation |
EN61326-1:2006 |
| Accessories |
Power cable, USB cable, Test lead, Fuse, CD-ROM (Application & Instruction Manual) |
- *1 : Specifications are for 2-hour warm-up condition, 10PLC and they are relative to calibration standards.
- *2 : 20% over range on all ranges except 1000VDC and 3A/10A range.
- *3 : Specifications are for 4-wire or 2-wire when a Null operation is used
- *4 : The range 750VAC is limited to 100kHz.
- *5 : Specifications are for sine wave input > 5% of range. (3Hz bandwidth)
- *6 : The Null function must be used.